Blue group
edited
... Standard Imaging BeamChecker Plus - http://www.standardimaging.com/product_home.php?id=13
Tol…
...
Standard Imaging BeamChecker Plus - http://www.standardimaging.com/product_home.php?id=13
Tolerances (expected): Dosimetry:Dosimetry:1
X-ray output constancy - 3%
Electron output constancy - 3%
...
Procedure / Tolerances
Daily quality assurance (QA) on a linear accelerator requires some basic steps before suggested specifications can be tested. First, on a Varian Clinac 2100C, the key must be turned to the on position. The Gantry is then turned to the upright 180o or straight down 0o location and temperature and pressures are checked. The temperature should read 40o, the gas should read 35 psi, and the pump pressure reading should be 80 psi. If any of the temperature or pressures are off the desired values QA is placed on hold and an engineer is contacted. After turning on the 4DTC computer, the multileaf collimation system (MLC) is turned on. Upon completion of testing all 60 MLC, the daily Linac QA may start.
...
and output tests.1tests.2
Safety checks:
a) Door interlock
...
{DSC01707.JPG}
References 1.Kutcher1. Khan FM. The Physics of Radiation Therapy. 4th ed. Philadelphia, PA: Lippincott Williams & Wilkins; 2010(398).
2.Kutcher G, Coia
Monthly Treatment Machine QA:
Brandon and Bret
Blue group
edited
... Merrill R, Miller M. Treatment planning system quality assurance. American Association of Medi…
...
Merrill R, Miller M. Treatment planning system quality assurance. American Association of Medical Dosimetrists Web site. http://medicaldosimetry.org/meetings/2012Presentations/061012/Merrill.pdf. Accessed Sep. 26, 2013.
Able CM. Treatment planning system QA in the era of image guided radiotherapy. Able Medical Physics Web site. http://www.ablemedicalphysics.com/uploads/TPS_QA_in_the_Era_of_IGRT.pdf. Accessed Sep. 26, 2013.
Blue group
edited
... {DSC01707.JPG}
References
... radiation oncology. Medical Physics. 1994; 21(4):587-592. …
...
{DSC01707.JPG}
References
...
radiation oncology. Medical Physics. 1994; 21(4):587-592.Med Phys. 1994;21(4):587-592.
Monthly Treatment Machine QA:
Brandon and Bret
...
Sun Nuclear Map Check
References:
...
Sun Nuclear Corporation. Available at:Corporation Web site. http://www.sunnuclear.com/home.asp. Accessed
...
2013.
2. IMRT QA II. The American
...
Physicists in Medicine. Available at:Medicine Web site. http://www.aapm.org/meetings/05AM/pdf/18-2601-49028-323.pdf. Accessed
Conventional Simulator QA:
Becky Holly Dustin
...
{IMG_1546.jpg}
References:
...
Simulator. Nucletron Website.Web site. http://www.nucletron.com/en/ProductsAndSolutions/Pages/SimulixEvolution.aspx
2013. Accessed September 25, 2013.
2. Khan FM. The Physics of Radiation Therapy. 4th ed. Philadelphia, PA: Lippincott, Williams and Wilkins; 2010.
...
{New Picture (7).png}
References:
...
quality assurance. American Association of Medical Dosimetrists Web site. http://medicaldosimetry.org/meetings/2012Presentations/061012/Merrill.pdf. Accessed
...
Able CM. Wake Forest University school of Medicine. Treatment planning
...
guided radiotherapy. Able Medical Physics Web site. http://www.ablemedicalphysics.com/uploads/TPS_QA_in_the_Era_of_IGRT.pdf. Accessed
Mijinheer B, Olszewska A, Fiorino C, et al. Quality assurance of treatment planning systems practical examples for non-IMRT photon beams. 1st Ed. Brussels, Belgium: ESTRO; 2004.
Blue group
edited
... 2. The American Association of Physicists in Medicine. Available at: http://www.aapm.org/meeti…
...
2. The American Association of Physicists in Medicine. Available at: http://www.aapm.org/meetings/05AM/pdf/18-2601-49028-323.pdf. Accessed October 2, 2013.
Conventional Simulator QA:
Becky Holly Dustin
Procedures:
{Nucletron Simplicity.JPG}
Blue group
edited
... {New Picture (4).png} Image of an electron density phantom used to test accuracy of CT data tr…
...
{New Picture (4).png} Image of an electron density phantom used to test accuracy of CT data transfer
{New Picture (1).png} Image of a Quasar body phantom used to test area and volumetric accuracy